Autoclave requalification problem
Problem: During routine empty chamber requal of the autoclave, we are
seeing a 4 degree C overshoot at start of exposure exceeding the HTM2010
limit of 3 degrees. This overshoot happened during heat up in previous
requals ie it appears to have shifted by up to 30 secs. Also seeing an
increased delta temp between the installed drain probes and chamber
probes of 0.8C versus 0.3C during previous quals and an increased delta T
of 3.08C from 1.5C across the validation probes. Also measure 6C
superheat in header before the autoclave control valve per HTM2010
method versus 0 - 0.7C before.
Checked all possible caused (we could think of):
1 Measurement - recalibrated all probes to same working standard and confirmed several times - ok.
We originally had a problem with the I/O card and repleced same and the
CPU card was replaced because the battery voltage was slightly low -
all control parameters checked to be ok. eproms not touched.
2 checked status of bleed valves and condensate bypass valve settings - ok
3
Jacket pressure switch dropped from 1.3 Bar to 0.8 Bar to eliminate
Jacket superheating the chamber. Temporary change as is likely to lead
to wet loads.
4 Control Valve operation - OK
5 PID values -
altered the settings to get a compliant cycle but it is not the root
cause and would likely lead to problems with distribution studies if
left.
6 CSG functioning correctly - removed some lagging after
PCV at CSG which is ~ 100m from Autoclave, to eliminate superheat &
to see if it had impact on cycle - no significant impact - still failed
cycle.
7 Cannot understand superheat at autoclave on CGS side of
PCV as there is nothing to generate it and the temp corresponds to sat
stm press in header. We use the HTM2010 method of measuring for
superheat. Reconfirmed 3 times. No measurable superheat in chamber using
packed towel test.
8 Need to find root cause - would really welcome any thought / suggestions
The superheat downstream of control valve results due to dryness in
steam. The approximate process across control valve is isenthalpic and
there is a temperature gain with the pressure reduction. You can find
out the amount of superheat if you know the pressures upstream and
downstream, from steam tables. If there is superheating, removing
insulation may not be effective.
It is better to generate pure steam at relatively low pressures.
Secondly, as pointed out by MedicineEng, air pockets(in chamber) may be a problem.
MORE NEWS